Setra-system Model 869 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Czujniki Setra-system Model 869. Setra System Model 869 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 28
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
SS869 8/12/08 Rev B
Apply power to calibrator with AC adapter or charge battery.
Turn on power switch located in the top right corner. Power is on when green light illuminates - the pocket PC will flash an orange light.
Push Pocket PC power button in top right of the unit. It will display a “start” screen. Tap “start” and select Setra Calibrator icon from “last
programs run” or “programs” file from drop down menu.
Calibrator starts in “Real Time” mode. Tap UUT (Unit Under Test) tab and select a stored UUT ID that corresponds to the pressure sensor
you wish to test - (tap the down arrow in the top box to view all of the stored UUT test profiles). Power is supplied to test pressure sensor
when UUT ID is selected. If a needed ID is not found, see Creating and Modifying UUT ID’s below.
Plug in pressure tubing and electrical cables into marked sockets on the top panel. Connect pressure tubing to “high” and “low” pressure
fittings. Hook up connections as follows:
1. 4-20mA current loop operation, connect +EXC to red 869 banana jack and -EXC to black jack.
2. For 3-wire voltage output operation, connect +EXC to red electrical jack, -EXC common to black jack, and +Output to green
jack.
Return to “Real Time” screen to verify correct zero pressure output of test pressure sensor. Tap “tare” button to re-zero onboard pressure
references (approx 10 seconds). You can also check full-scale output by tapping “+FS” (+ full scale) button. You can also make zero and
span adjustments to test pressure sensor at this time.
Go to “Test” screen and tap “Do Test” button to initiate automatic calibration test. The calibrator will apply test pressures and record output
results until “test complete” flag is displayed. You can save the test data by tapping the “Save Test Results” button. The data is saved with
Micro-Cal
®
Model 869 Quick Start Guide
To create a new UUT ID go to the UUT screen and tap “new” button. Next, type in a new UUT ID in the upper box. Enter data by tapping
the keyboard icon in lower right of screen. Characters are entered on screen at the cursor when mini-keyboard characters are tapped. An
existing ID template must be selected in “copy from” box. It is suggested that an existing ID template be chosen that closely matches the
pressure sensor desired to test either in pressure range or output type. Tap “ok”. When screen returns to UUT, any ID values can be modi-
fied using keypad. Use dropdown boxes to change excitation, output type, or number of test points. Hit “save” button and new test profile is
ready to use.
To modify a UUT ID, select desired UUT ID from dropdown screen, make changes, and tap “save”. The newly created or modified UUT IDs
will be available from dropdown UUT ID screen.
Creating and Modifying UUT ID’s
I
II
III
IV
V
VI
VII
Setra Systems, Inc., 159 Swanson Rd, Boxborough, MA, 01719
User Manual
Micro-Cal™
Model 869
Very Low Pressure Generating and Documenting Calibrator
800.257.3872 • Toll Free
978.264.0292 • Fax
setra.com • Web Site
pressure transducers
Software Version V3.02
SS869 2/03/12 Rev. D
Software Version V3.21
Przeglądanie stron 0
1 2 3 4 5 6 ... 27 28

Podsumowanie treści

Strona 1 - Model 869

SS869 8/12/08 Rev BApply power to calibrator with AC adapter or charge battery.Turn on power switch located in the top right corner. Power is on when

Strona 2

10 5.3 USER INTERFACE PANEL The user interface panel includes the pressure and electrical connections to the unit under test. (See Diagrams 5A &am

Strona 3 - Table of Contents

46 11 5.4 GETTING STARTED 1. Open calibrator top. 2. Turn on calibrator power (power switch on right side of the calibrator). 3. Once calibrator

Strona 4

126.0 POCKET PC INTERFACE SCREENS The Pocket PC Interface is tab-based, split into 6 major sections at the bottom of the screen. 6.1 APPLICATION DE

Strona 5 - 1.0 SAFETY INSTRUCTIONS

4412.0 SPECIFICATION DERIVATIONSTotal Error Band(Worst case) over atemperature range of-10°F to +130°FNon Linearity: ±0.1% FSHysteresis: ±0.05% FSNon

Strona 6 - Diagram 1

14 6.3 TEST: General Screen Instructions (Diagram 8): 1. Tap the TEST tab at the bottom of the page 2. Select desired UUT ID from the drop down b

Strona 7 - Diagram 2

4210.0 GLOSSARY OF TERMS Atmospheric Pressure - Pressure of the atmosphere at the Earth’s surface. NIST standard atmospheric pressure = 1.01325 bar

Strona 8 - Battery Jack

16 6.33 TEST: Dither Control Page (Manual Pressure Incrementing) Screen Instructions (Diagram 10): Test Point Indicator: Displays present test poi

Strona 9 - 5.0 CALIBRATOR SETUP

40 Once a UUT ID is selected, the test results list will update to display only the test results from that UUT. (See Diagram 36) The test results fo

Strona 10 - Non-Linearity

18 6.4 PROFILE: General Setting up Test Profiles Database (Diagram 11): 1. Profiles are pressure settings used during testing. 2. Program has cap

Strona 11 - Diagram 6

38 New Button - This button opens a dialog box to create a unique profile. Profile ID must be entered into the ID box on the dialog. 1. To copy a pr

Strona 13 - Diagram 7

20 6.5 UUT (UNIT UNDER TEST) SETUP: General Screen Instructions (Diagram 13): UUT setup screeen is used for inputting UUT profiles into the UUT da

Strona 14 - 11.0 SPECIFICATIONS

36 Pocket PC Database - This list shows what test profiles are currently on the Pocket PC. After editing any profile, said profile will automaticall

Strona 15 - Diagram 9

22 UUT SUPPLEMENTAL PAGE 1. Enter the UUT Part Number. 2. Enter the UUT Model. 3. Enter the UUT Manufacturer. 4. Enter the required pressure Settlin

Strona 16 - Diagram 39

349.0 MICRO-CAL MANAGER INTERFACE The MICRO-CAL MANAGER INTERFACE provides a central location for all data within a given facility, database manageme

Strona 17 - Diagram 37

24 6.51 UUT (UNIT UNDER TEST) SETUP: Dial Gauge Differences from General (Diagram 13): -Output fields hidden. -Excitation field hidden. 6.52 UUT

Strona 18 - Diagram 35

32Electrical LinesUnit Under Test(Model 260)Pneumatic LinesDiagram 22_ + EXC24VDCSIG(+)SIG(-)COMCurrent InternalExcitationEXC COM SIG

Strona 19

26 6.7 SYSTEM Screen Instructions (Diagram 17): Data and system values: Software Version - Version of software on Pocket PC. Firmware Version -

Strona 20 - Diagram 31

27Diagram 17Default System SettingEng. Units SettingSoftware VersionFirmware VersionSerial NumberFilter LengthExcitationBattery LevelLeak Test System

Strona 21 - Diagram 13

28 EDIT DEFAULTS PAGE Display trend graph on UUT warm-up box will make a graph appear when the DoTest Button is pressed on test page.The graph will

Strona 22 - Diagram 27

3Table of Contents1.0 SAFETY INSTRUCTIONS ... 52.0 INTRODUCTION ...

Strona 23 - Diagram 14

49.0 MICRO-CAL MANAGER INTERFACE 9.1 Setup for Pocket PC Connection ... 34 9.2 Graphical User Interface Setup ...

Strona 24

51.0 SAFETY INSTRUCTIONS Use the calibrator only as specified in the manual, otherwise the protec-tion provided may be impaired. Do not operate

Strona 25 - Diagram 16

62.0 INTRODUCTION DOCUMENTING PROCESS CALIBRATOR: The Setra Micro-Cal Model 869 is a pressure generating, very low pres-sure documenting calibrator

Strona 26 - 6.7 SYSTEM

50 73.0 STANDARD EQUIPMENT BREAKOUTTHE ITEMS BELOW ARE INCLUDED IN YOUR CALIBRATOR:Diagram 2AC Power AdapterPneumatic Fittings & TubingElectronic

Strona 27 - Diagram 20

84.0 OPERATING FEATURES 4.1 THE POCKET PC PERFORMS THE FOLLOWING MAJOR FUNCTIONS • User interface for setups, display and storage. • Data trans

Strona 28 - Restore Factory Defaults

4813.0 SPARE PARTS & ACCESSORIES Part # Item 269701 269 terminal block 269425-02 security key 869902 accessory bag 869904-xx elec

Komentarze do niniejszej Instrukcji

Brak uwag