Setra-system Model 201 Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Czujniki Setra-system Model 201. Setra System Model 201 User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 4
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
Setra Systems Model 201
Pressure Transducers
1.0 GENERAL INFORMATION
Every Model 201 has been tested and calibrated before shipment. Specific
performance specifications are shown on page 3 of this Guide.
Setra Systems 201 pressure transducers sense gauge and differential pressure and
convert this pressure difference to a proportional high level analog output. The
Model 201 is a 4 to 20 MA output device.
2.0 MECHANICAL INSTALLATION
2.1 Media Compatibility
Model 201 transducers are designed to be used with any gases or liquids
compatible with 316 Stainless Steel and Inconel® in process port. Reference port
must be exposed to clean dry air or non-corrosive gas.
2.2 Environment
The operating temperature limits of the 201 are -40° to +175°F (-40 to
+80°C). The compensated temperature range is -25 to +175°F (-33 to +80°C).
2.3 Pressure Fittings
Typically, standard pipe fittings and procedures should be used. Excessive torquing
of metal fittings may cause a slight zero shift. The use of plastic fittings typically
results in no noticeable zero shift. Torquing does not appreciably affect linearity or
sensitivity.
2.4 Venting for Gauge Pressure Measurements
All 201 units are vented to atmosphere through a Barbed reference port at the
bottom of the unit.
2.5 Differential Pressure Measurements
Connect low pressure tube to reference port. Note: Media into reference port must be
clean dry air or non-corrosive gas.
Installation Guide
CONDUIT
FITTING
1/2 NPT
7/8 HEX
1/4 - 18 NPT MALE
1.96
50
POSITIVE PRESSURE
PROCESS PORT
5.41
137
2.00 DIA
50 DIA
REFERENCE PORT
BARBED FITTING
FOR Flexible 1/4” ID
TUBE
CONNECTION
Przeglądanie stron 0
1 2 3 4

Podsumowanie treści

Strona 1 - Installation Guide

Setra Systems Model 201Pressure Transducers1.0 GENERAL INFORMATIONEvery Model 201 has been tested and calibrated before shipment. Specificperformance

Strona 2 - 3.0 ELECTRICAL INSTALLATION

23.1 Model 201 OutputThe Model 201 (current output) is a true 2-wire, 4-20 mA current output device thatdelivers rated current into any external load

Strona 3

34.0 CalibrationAll Setra current pressure transducers are carefully calibrated to the specific inputpressure range vs. output current at the factory

Strona 4 - SS2118 Rev.A 04/05/05

4159 Swanson Road, Boxborough, MA 01719-1304Tel: 800-257-3872/978-263-1400 SS2118 Rev.A 04/05/055.0 RETURNING PRODUCTS FOR REPAIRBefore returning th